next up previous contents index
Next: 7.6 Wavelength calibration accuracy Up: 7 SWS Calibration Previous: 7.4 Flux calibration accuracy

7.5 Wavelength calibration

 

The astronomical quantity of interest is an array of flux as a function of wavelength. Hence, a relationship must be established between each grating scanner position (measured by a Linear Voltage Differential Transducer (LVDT)  and colloquially given the units `LVDT') as contained in the ERD and its corresponding wavelength. The relation is calculated in two steps: first the scanner position is converted to a scanner angle using a calibration file. Then the following equation can be applied to calculate the wavelength of light falling on the detectors:

  equation1180

where,

tex2html_wrap_inline7403
is the angle from the optical axis  to the entrance or light source used, as seen from the collimator,
tex2html_wrap_inline7585
is the angle from the optical axis to the detector pixel,
tex2html_wrap_inline7587
is the angle of incidence on the grating for a beam entering along the optical axis for the current scanner position,
D
is the grating constant ,
N
is the spectral order .

tex2html_wrap_inline7593 and D are constants, taken from calibration file 16 . The spectral order N is determined by calculating all the orders that give wavelengths consistent with the transmission limits of the band/order/aperture/filter combination. For valid observations only one order should give a wavelength that satisfies this criterion.

The wavelength range - band/order/aperture/filter combination relationship is shown in Table 7.3.

For the Fabry-Pérot , first the grating wavelength is calculated. Calibration file 12  then transforms the nominal FP-position  (an engineering parameter of no interest to the observer) into a FP-gap . This mechanical gap width must be corrected to get the optical gap width. The wavelength which passes through the FP is the one of which a whole multiple fits exactly in two times the optical gap width.

 

band grating entr. det. Wavelength
order apert. array range
1A S4 1 1 2.38 - 2.61
1B S3 1 1 2.60 - 3.03
SW 1D S3 2 1 3.02 - 3.53
section 1E S2 2 1 3.52 - 4.08
2A S2 2 2 4.05 - 5.31
2B S1 2 2 5.30 - 7.01
2C S1 3 2 7.00 - 12.1
3A L2 1 3 12.0 - 16.6
LW 3C L2 2 3 16.5 - 19.6
section 3D L1 2 3 19.5 - 27.6
3E L1 3 3 27.5 - 29.0
4 L1 3 4 28.9 - 45.2
FP 5A L3 1 5 11.4 - 12.2
5B L2 1 5 12.2 - 16.0
5C L2 2 5 16.0 - 19.0
5D L1 2 5 19.0 - 26.0
6 L1 4 6 26.0 - 44.5
Table 7.3: SWS band wavelength parameters

 


next up previous contents index
Next: 7.6 Wavelength calibration accuracy Up: 7 SWS Calibration Previous: 7.4 Flux calibration accuracy

K. Leech with contributions from
the SWS Instrument Dedicated Team (SIDT)
and the SWS Instrument Support Team (SIST)