Description of CAM03



next up previous contents
Next: Input Parameters of Up: CAM03 - BEAM-SWITCHING Previous: CAM03 - BEAM-SWITCHING

Description of CAM03

 

CAM03 uses a spatial modulation: the background is measured with the same pixel, first on the source and next on a nearby empty reference field. Assuming that it has the same intensity in both observations, it can be accurately subtracted, regardless of the flat-field accuracy. The limitation of this technique is the small scale spatial structure of the background, which is not known at this time. Extrapolation towards high spatial frequency of the power spectrum of these fluctuations in the IRAS data indicates that this sensitivity limitation should be at a very low flux level.

Using only one reference is the normal case, but CAM03 allows to take 1 to 4 different reference fields. This possibility allows to average out a large scale gradient in the background.

Each passage on a field is a pointing. The same number of exposures is taken at each pointing, whether source or reference field, so that the total time spent on the source is the same as the total time spent observing reference fields. The fields are observed in the same sequence, defined by the observer, for all the observations within an AOT CAM03: Source Ref1 Source Ref2 etc.

This sequence defines a cycle of pointings. In the absence of low frequency noises (drift etc.), the observation can be completed in one cycle. Nevertheless, if is large, it may be better to split the observation in several cycles, at some slew time overhead expense. The observer should limit the total duration of a cycle to 20 minutes, and split the observation in several cycles if this limit is exceeded.

A cycle of pointing is always performed in the same optical configuration, but, within a single CAM03, always using the same set of fields, it is possible to make observations with up to 64 different optical configurations (channels, filter or CVF , PFOV).

Raster scans, CVF scans, polarimetric observations, and the FLASH-MODE of the SW channel are not supported in CAM03. Still, the CVF can be used as a narrow band pass filter as in CAM01.

The total time spent for an observation is the sum of the 3 terms:

which are described in sections gif, gif and gif



next up previous contents
Next: Input Parameters of Up: CAM03 - BEAM-SWITCHING Previous: CAM03 - BEAM-SWITCHING



ISOCAM Observer's Manual - V1.0
Tue Oct 31 12:06:23 MET 1995